Standard Specification for Thermocouple Materials, Platinum-Rhodium Alloys, and Platinum (Includes all amendments and changes 5/14/2021).
Translate name
STANDARD published on 1.11.2020
Designation standards: ASTM E1159-15(2020)e1
Publication date standards: 1.11.2020
SKU: NS-1013211
The number of pages: 3
Approximate weight : 9 g (0.02 lbs)
Country: American technical standard
Category: Technical standards ASTM
Temperature-measuring instrumentsOther non-ferrous metals and their alloys
This specification covers non-insulated platinum rhodium alloys (weight percent composition), and platinum thermo elements. The purity of platinum for use in the negative thermo element of Type R and Type S thermocouples shall be of a specific Grade, as specified. The purity of platinum before alloying for use in the positive thermo element of Type R, S, and both thermo elements of Type B shall also be of a certain grade, as specified. All material to be supplied shall consist of continuous lengths which shall be free of kinks and shall be round and smooth. Any pits or dents that exceed a certain percentage of the diameter shall be cause for rejection of the material. After material has been processed to its final diameter and prior to subsequent anneal, it shall be cleaned to remove all dirt, lubricant, fingerprints, or other foreign residue. The thermoelements shall be such that, when matched as a thermocouple, the EMF output shall meet the tolerance on the initial value of EMF versus temperature as specified.
Keywords:
platinum, rhodium, thermocouple materials, thermoelement, type B, type R, type S,, ICS Number Code 17.200.20 (Temperature-measuring instruments),77.120.99 (Other non-ferrous metals and their alloys)
1. Scope | ||||||||||||
1.1?This specification covers non-insulated platinum-rhodium alloys (weight percent composition), and platinum thermoelements that meet the requirement of Specification E230/E230M and NIST Monograph 175.1.2?This specification does not cover platinum and platinum-rhodium materials, that require a higher purity than specified in 5.1, such as used for temperature coefficient of resistance (TCR) measurements or standards type calibration. For requirements of this superior quality, it is suggested that suppliers of precious metals be contacted. 1.3?The values stated in SI units are to be regarded as standard. The values given in parentheses are mathematical conversions to inch-pound units that are provided for information only and are not considered standard. 1.4?This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.5?This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee. |
||||||||||||
2. Referenced Documents | ||||||||||||
|
Do you want to make sure you use only the valid technical standards?
We can offer you a solution which will provide you a monthly overview concerning the updating of standards which you use.
Would you like to know more? Look at this page.
Latest update: 2024-12-21 (Number of items: 2 216 840)
© Copyright 2024 NORMSERVIS s.r.o.