Standard Specification for Polyethylene Plastics for Medical Applications
STANDARD published on 1.8.2009
Designation standards: ASTM F639-09
Note: WITHDRAWN
Publication date standards: 1.8.2009
SKU: NS-55977
The number of pages: 3
Approximate weight : 9 g (0.02 lbs)
Country: American technical standard
Category: Technical standards ASTM
Keywords:
plastic surgical devices/applications, polyethylene (PE) plastics´surgical implant applications, polymers´surgical applications, Plastic surgical devices/applications--specifications, Polyethylene (PE) surgical implant applications--specifications, Polymers (surgical applications)--specifications, ICS Number Code 11.120.01 (Pharmaceutics in general), 83.080.20 (Thermoplastic materials)
Significance and Use | ||||||||||||||||||||||||||
SignificanceTop
X1.2.1 Concentrations of trace metals are measured as extracts in simulated body fluids. The metal’s concentration in extracts is based on the surface area of the plastic extracted from which the total amount of metal deliverable to the patient may be estimated. |
||||||||||||||||||||||||||
1. Scope | ||||||||||||||||||||||||||
1.1 This specification covers polyethylene plastics (as defined in Terminology D 883) intended for use in medical device applications involving human tissue contact devices, short-term indwellings of 30 days or less, and fluid transfer devices. The biocompatibility of these materials as a class has not been established. Biocompatibility tests must be conducted on the final product. 1.2 This specification is not applicable to ultra-high molecular weight polyethylenes (UHMWPE) plastics, such as those used in joint implants, and so forth. 1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. |
||||||||||||||||||||||||||
2. Referenced Documents | ||||||||||||||||||||||||||
|