Standard Practice for Evaluating Acute Systemic Toxicity of Material Extracts by Systemic Injection in the Mouse
Translate name
STANDARD published on 1.6.2020
Designation standards: ASTM F750-20
Publication date standards: 1.6.2020
SKU: NS-1001428
The number of pages: 4
Approximate weight : 12 g (0.03 lbs)
Country: American technical standard
Category: Technical standards ASTM
Keywords:
acute toxicity tests, biocompatibility, intravenous injection, intraperitoneal injections, mouse/mice, test animals ,, ICS Number Code 11.220 (Veterinary medicine)
Significance and Use | ||||
4.1 This practice is intended to help assess the biocompatibility of materials used in medical devices. It is an acute toxicological test designed to detect the presence of injurious leachable substances. 4.2 This practice may not be appropriate for all types of implant applications. The user is cautioned to consider the appropriateness of the method in view of the materials being tested, their potential applications, and the recommendations contained in Practice F748. 4.3 The only limitation applicable is the extract preparation. Refer to Sections 4.3 and 4.4 of Practice F619 for a description of this limitation. |
||||
1. Scope | ||||
1.1 This practice covers a nonspecific, acute toxicity test used for detecting leachables from materials used in medical devices. 1.2 The liquids injected into the mouse are those obtained by Practice F619 where the extraction vehicles are saline, vegetable oil, or other liquids simulating human body fluids. 1.3 Two procedures are outlined: Method A for intravenous injection and Method B for intraperitoneal injection. 1.4 This practice is one of several developed for the assessment of the biocompatibility of materials. Practice F748 may provide guidance for the selection of appropriate methods for testing materials for a specific application. 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee. |
||||
2. Referenced Documents | ||||
|