Dissolved Molecular Oxygen in Liquid Hydrocarbons by Electrochemical Detection
STANDARD published on 25.7.2002
Designation standards: UOP 678-88
Note: WITHDRAWN
Publication date standards: 25.7.2002
SKU: NS-574853
The number of pages: 5
Approximate weight : 15 g (0.03 lbs)
Country: American technical standard
Category: Technical standards UOP
1. Scope |
This method is for estimating the dissolved oxygen in liquid hydrocarbons by electrochemical detection. The method is applicable to both on-line and laboratory analyses. The lower detection limit is 0.1 mass-ppm while the upper limit is dependent on the solubility of oxygen in the specific sample. Carbon dioxide, ammonia and hydrogen sulfide at concentrations above 200, 300 and 15-ppm (see Note 1), respectively, will interfere with the oxygen determination. Materials that contaminate the membrane or volatile components that are either oxidized, reduced, or react with the electrodes will also interfere. The sampling system associated with a UOP process unit will be specified by UOP depending on the specific process unit. |